The QuickTOCpurity by LAR Process Analysers is a customised analyser of total organic carbon content for pure and ultrapure water applications. The sample is injected via an injection loop, avoiding environmental influences to the sample. In the oven, the thermal oxidation at 1,200 °C takes place. Then, the produced CO2 will be measured by a NDIR detector resulting in the exact determination of TOC. The system is able to measure up to six sample streams in a unit. It is operated by a self-explanatory software via a 10.4” touchscreen display. Due to the integrated calibration and validation technique, that uses a special test gas, operators have the opportunity to check the analyser automatically and remotely at any time.