Multi-component analyser system
The multi-component analyser system MCS300P from Sick provides reliable measurement results thanks to photometric process gas monitoring. In the MCS300P HW system for process monitoring in exhaust gas purification systems, smoke gas is continuously sampled via a sampling probe with a heated filter before being fed through a heated measuring gas line to the analyser, which is itself housed within a heated gas cell. Its original composition including water content is retained. As a result, there is no risk of corrosion due to condensation. The MCS300P gas analyser sends its signals to the process controller. Thanks to the combination of a non-dispersive, photometric measurement procedure and flexible measuring cuvettes, all active components in the infrared spectrum and visual spectrum can be measured, from the most minute amounts (ppm) up to large volume concentrations (vol%). The pressurised MCS300P-Ex variant can be used in Ex zones 1 and 2 as a 3G or 2G ATEX class device. Up to six components can be identified simultaneously using the double filter wheels.
www.cpp-net.com search: cpp0315sick